NGT-CN High Sensitivity Gamma Tool

NGT-CN

High Sensitivity Non-Directional Gamma Tool
(1.36” OD)

The NGT-CN High Sensitivity Non-Focused Gamma Tool is similar to the NGT-T model but has 40% higher sensitivity due to its longer scintillator crystal. The NGT-CN is interchangeable with the standard NGT-T gamma tool. It is the same overall length and diameter as the NGT-T but is used in high sensitivity non-focused gamma applications.

To protect your gamma investment and ensure the best performance use CBG original SNB Snubbers-Shock & Vibration Isolators.

PN 201-000-000_NGT-CN-800x503
PN 201-000-000_NGT-CN-800x503

NGT-CN

High Sensitivity Non-Directional Gamma Tool
(1.36” OD)

The NGT-CN High Sensitivity Non-Focused Gamma Tool is similar to the NGT-T model but has 40% higher sensitivity due to its longer scintillator crystal. The NGT-CN is interchangeable with the standard NGT-T gamma tool. It is the same overall length and diameter as the NGT-T but is used in high sensitivity non-focused gamma applications.

To protect your gamma investment and ensure the best performance use CBG original SNB Snubbers-Shock & Vibration Isolators.

NGT-CN Specifications

P/N 201-000-000

Detector Scintillator Crystal Sodium Iodide, Na(TI)
Photomultiplier R3991A (Hamamatsu)
Mechanical Diameter (OD) 1.36”
Length (make-up) 13.6”
Weight 1.9 lb.
End Connectors Micro-D Subminiature, 15-Pin
Material Aluminum Alloy
Performance Sensitivity 1.59 CPS/API*
Accuracy +/- 5% to 150°C
+/- 10% to 175°C
Environmental Operating Temperature -18°C to 175°C
Survival Temperature (reduced accuracy) -40°C to 200°C
Heat/Cool Rate, Max. 2.8°C (5°F) per minute
Vibration (3 axis) 50-1000Hz, Random 30G (20G RMS)
Shock (3 axis) 1000G, 0.5ms, half-sine
Power Requirements DC Input Voltage 17-48 VDC, 50VDC Max.
Tool Current 8-11 mA
Output Signal Negative Pulse +5V to GND, 5 uSec width

* Counts are collected with gamma tool mounted within a standard BeCu pressure barrel, (OD 1.875″, ID 1.500″).  API Calibration performed with AEA Technology, QSA Inc., KUTh Field Verifier #188074, S/N 255 Radionuclides: K-40(11.7%), U-238(90 ppm), Th-232(40 ppm).